Products

IRTP3-150D

1. General information

Compact pulse not protected from lost radiation X-ray tube IRTP3-D with cold cathode is designed for medical X-ray treatment of the sick in stationary and nonstationary conditions and for industrial X-raying materials. The tube can be applied in local fault detector “Arina-1”, “Arion” and in X-ray machinery under development.

2. Basic specifications

2.1 Electrical and X-ray specification

Characteristic, unit measure

rate

test data

Appendix

min

nominal

max

Tube voltage, kV

90

 

150

 

 

X-ray radiation exposure doze for a pulse (average value), mrad

25

 

 

 

1,2

The size of efficient focus spot (diameter), mm

 

1,5

1,8

 

 

Notice:
1.With the tube voltage 150 kV, pulse rate 15 s-1 and capacitance discharged through the tube, C=30 petapharad.
2.The period of delay 1s, the amount of switching-on 10, the distance from the tube window 50 mm.
2.2  Marginal operating conditions
Notice:
1.With the tube voltage 150 kV, pulse rate 15 s-1 and capacitance discharged through the tube, C=30 petapharad.
2.The period of delay 1s, the amount of switching-on 10, the distance from the tube window 50 mm.
2.2  Marginal operating conditions

Characteristic, unit measure

rate

min

max

Tube voltage, kV

90

150

Pulse rate, Hz

 

15

Capacity discharge on the tube, pf

 

30

Cycling operating mode: tube operating time for an hour with time rate 15Hz, min

 

15

Notice: the operating of the tube is acceptable with the voltage far exceeding 150 kV but in case of reducing current period through the tube (chopping gap) about several units ns.
2.3 Minimal lifetime – 106 pulses.
 

After-Sales

The goods are custom-made.
Please, call us and agree on the delivery time with our manager.
+7 4912 24-90-92